CALIBRATION OF TLD IN THE CARBON ION BEAM RBS U-70
Abstract
The results of TLD calibration in the RBS U-70 carbon beam are presented. Irradiation was performed at
different depths of the water phantom. The LET value at the place of irradiation was estimated from the results of
calculations using the MCNPX program. The LET range was from 12 to 200 keV/μm. A method is proposed for
the insertion of a correction for TLD readings according to the dependence of the high-temperature ratio in the
thermoluminescence curve on the LET value.
References
2. Postanovlenie Pravitel'stva Rossijskoj Federacii ot 16.03.2019 №287 «Ob utverzhdenii Federal'noj nauchno-tehnicheskoj programmy razvitija sinhrotronnyh i nejtronnyh issledovanij i issledovatel'skoj infrastruktury na 2019 - 2027 gody»
3. Alekseev A.G., Kirjuhin O.V., Batuhtina O./ Issledovanie zavisimosti harakteristik termoljuminescentnyh detektorov ot velichiny LPJe izluchenija/ Zhurnal: Evrazijskij Sojuz Uchenyh (ESU) # 2 (59), 2019. https://www.doi.org/10.31618/ESU.2413-9335.2019.1.59.43-47 https://euroasia- cience.ru/wpcontent/uploads/2019/03/43-47-Alekseev-A.G.-Kiryukhin-OV-Batukhtina-OI.pdf. (data obrashhenija - 24.09.2022g.).
4. Pikalov, V.A. et al., “Experimental Facility "Radiobiological Test Setup on Accelerator U-70" as Centers for Collective Use (CCU)”, in Proc. RuPAC2021, Alushta, Russia, 2021. P. 124-126. doi:10.18429/JACoW-RuPAC2021-FRB06.
5. Sozdanie specializirovannogo grebenchatogo fil'tra dlja poluchenija ravnomernoj biologicheskoj dozy na puchke ionov ugleroda / Solov'ev A.N. [i dr.] // Medicinskaja fizika. 2021. T. 89. № 1. S. 48.
6. N. Vana, W. Schöner, Y. Akatov / Absorbed Dose Measurement and LET Determination with TLDs in Space Radiation Protection Dosimetry, Volume 66, Issue 1-4, 1 July 1996, Pages 145–152, https://doi.org/10.1093/oxfordjournals.rpd.a031703
7. Radiation Medicine QA Solutions Edition Nov 2019 https://www.ptwdosimetry.com/en/support/downloads/?type=3451&downloadfile=1816&cHash=fdec9ca50
388aa66ba76e25663a75687 (data obrashhenija - 27.09.2022g)
8. UNIDOS webline High quality Reference Class Dosemeter for radiation therapy, diagnostic radiology and health physics https://www.ptwdosimetry.com/en/support/downloads/?type=3451&downloadfile=1593&cHash=05709dfeb e4293b4223df8dd3955a0b8 (data obrashhenija - 27.09.2022g)
9. https://mcnpx.lanl.gov/; .S. Hendricks, et al., "MCNPX 2.6.0 Extensions", LA-UR-08-2216 (2008). https://mcnpx.lanl.gov/reference_collection.html#mcnpx_refs
10. E Haettner E[herimental Study jn Carbon Ion in Water Using GSI Therapy Beams [master’s thesis] https://www.kth.se/polopoly_fs/1.144271!/Menu/gene ral/column-content/attachment/emmahaettner.pdf ( data obrashhenija18.02.2018g.).
CC BY-ND
A work licensed in this way allows the following:
1. The freedom to use and perform the work: The licensee must be allowed to make any use, private or public, of the work.
2. The freedom to study the work and apply the information: The licensee must be allowed to examine the work and to use the knowledge gained from the work in any way. The license may not, for example, restrict "reverse engineering."
2. The freedom to redistribute copies: Copies may be sold, swapped or given away for free, in the same form as the original.